−60 °C vs −80 °C freezers: which temperature do your samples need?

'Ultra-low' is often used as shorthand for −80 °C, but freezers that hold −60 °C and thereabouts serve plenty of real needs too. The difference is not just 20 degrees: it changes the refrigeration system, the energy bill, the room you need and how much margin you have when something goes wrong. This guide sets out the trade-offs so you can match the cabinet to the product. The storage temperature itself is never a buying decision — it comes from the product label, the test method, the sponsor or your institution's protocol.

7 min readUpdated

Start with the required storage temperature

Read the label, kit insert, study protocol or SOP. If it says 'store at or below −70 °C' or gives a range such as the −90 to −60 °C ultra-cold band described in US CDC vaccine guidance, a −60 °C freezer is not enough, whatever the price difference. If it says −20 °C or −40 °C, an ultra-low cabinet may be wasted energy.

Where the requirement is 'frozen, below −50 °C' or similar, a −60 °C class freezer may be a sensible fit. Confirm with whoever owns the requirement before buying.

How the two classes compare

General differences between −60 °C and −80 °C class freezers (models vary)
Aspect−60 °C class−80 °C class
Typical refrigerationSingle-stage or two-stage, depending on designUsually two-stage cascade or specialised systems
Energy useLower for the same volumeHigher; the gap to room temperature is larger
Heat into the roomModerateHigh — room ventilation or air-conditioning matters
Pull-down and recoveryGenerally quicker to reach set pointSlower after loading or door openings
FormatOften chest styleUpright and chest
Common usesLong-hold frozen goods, some reagents and samples rated for this rangeBiobanks, long-term biological samples, products labelled for ultra-cold storage

Why some samples need colder

Lower temperatures slow chemical and biological change. For some materials, the slower rate at −80 °C over many years of storage is what preserves them; for others, −20 °C or −60 °C is already enough. Reviews of lab practice have noted that direct comparisons of sample stability at −70 and −80 °C are limited, which is why the decision must rest on product data or protocol, not habit.

Margin also matters. A sample that must stay below −60 °C has more buffer in a −80 °C freezer during a door opening or a short power cut than in a freezer running right at −60 °C.

Energy and room considerations

Each degree colder costs energy. MIT's lab guidance says moving ULT freezers from −80 to −70 °C can save up to 30%; the step to a −60 °C class freezer is larger again. In India, where many labs run without round-the-clock air-conditioning, the heat a −80 °C unit puts into the room also affects the next freezer and the people working there.

Check the rated ambient for both options. A freezer that holds −80 °C in a 25 °C room may not hold it in a 35 °C one.

Chest or upright

  • Chest freezers hold cold air when opened and usually hold temperature longer after a power cut.
  • Upright freezers save floor space and make racked inventory easier to reach, at the cost of more cold air lost per opening.
  • For long-hold storage with infrequent access, chest designs are often more forgiving.
  • For daily-access sample storage, uprights with inner doors limit the cold lost per opening.

A quick decision guide

  • Product or protocol says ≤ −70 °C or −80 °C: choose a −80 °C class freezer.
  • Requirement allows −60 °C or warmer and access is infrequent: a −60 °C class freezer can cut energy and cost.
  • Mixed requirements: store each product in a freezer that meets its own requirement, rather than running everything at the coldest setting.
  • Unsure: ask the product owner or supplier for written storage conditions before buying.

Where KALT fits

KALT's deep freezer family covers −18 °C down to −60 °C. That suits applications whose requirement sits within that range; it is not a substitute for a −80 °C freezer where the product or protocol needs colder. Ask us for the specification of a particular model and compare it against your written requirement.

Frequently asked questions

Can I run a −80 °C freezer at −60 °C to save energy?
Often yes, and it will use less energy, but only if every product inside is rated for that temperature. Check labels and protocols first, and record the change in your SOPs.
Is −60 °C cold enough for mRNA vaccines?
Requirements differ by product. Some are labelled for ultra-cold storage in the −90 to −60 °C range; others have different conditions. Always follow the product's own storage statement and your programme's guidance.