First, check these
- Do not discard the vaccines and do not use them. Mark them clearly as 'do not use' and keep them together.
- Move them to another unit that is holding +2 to +8 °C, or to a validated cold box as your procedures describe.
- Record the current temperature, the minimum and maximum readings, the time the excursion was found and how long it lasted if your logger shows it.
- Report immediately to your supervisor or cold chain officer and follow your programme's excursion procedure.
- Check indicators such as vaccine vial monitors if your programme uses them, and record what you see.
Likely causes
1. Door left open or not sealing
You can check thisSigns: Warm excursion during busy sessions or after stock checks; door found ajar; gasket damaged.
Fix: Close the door and check the seal. Replace damaged gaskets and fit a door alarm if available. Train staff to close the door after each use.
2. Power cut
You can check thisSigns: Excursion lines up with an outage, logger or controller shows a power-fail event.
Fix: Keep the door closed during outages and follow your contingency plan. Backup power or units designed to hold temperature during outages are worth considering.
3. Vials too close to the cold source
You can check thisSigns: Cold excursion or frozen vials near the back wall, the evaporator outlet or the bottom of the unit.
Fix: Store vaccines in the zones your procedures specify, away from cold walls and air outlets, with space for air to circulate.
4. Setpoint changed or controller fault
Technician jobSigns: Temperature consistently high or low, setpoint different from normal, display disagreeing with a calibrated logger.
Fix: Restore the correct setpoint if your procedures allow and inform your supervisor. A technician should check the controller and sensor.
5. Unit in a hot room or near heat
You can check thisSigns: Warm excursions on hot afternoons, unit near a window or without ventilation.
Fix: Relocate to a cooler, ventilated position with proper clearances, and clean the condenser with power isolated.
6. Refrigeration fault
Technician jobSigns: Temperature drifting despite a closed door and stable power, compressor or fan not running.
Fix: Call a technician and move all stock to working storage.
When to call a technician
Call a technician for any excursion caused by an equipment fault, and keep the affected unit out of vaccine use until it has been repaired and has shown stable temperatures on a logger. Domestic refrigerators are not suitable for vaccine storage; use purpose-built medical refrigeration that meets your programme's requirements. The decision on the vaccines themselves always follows your programme rules.
Isolate power before cleaning or moving a refrigerator. Do not attempt repairs yourself. If the unit uses hydrocarbon refrigerant (R290 or R600a), only trained technicians should work on it.